Lesson 8 of 12 · 20 min
Flame tests and atomic absorption spectroscopy
Identify metal ions from flame colours, explain flame tests and AAS with energy levels, and use an AAS calibration curve, with any dilution, to find the concentration of a metal ion in solution.
What you will be able to do
- Carry out and interpret a flame test, and state its limitations.
- Explain flame tests (emission) and AAS (absorption) in terms of electrons moving between energy levels.
- Use an AAS calibration curve to find the concentration of a metal ion, including a dilution factor.
- Evaluate AAS data, including blanks, interpolation and uncertainty.
Fireworks get their colours from metal salts: strontium for red, barium for green, sodium for yellow. The same effect is a quick laboratory test for metal ions, and its absorption counterpart is one of the most widely used instruments in water testing, mining and medicine.
Key idea
How a flame test works
- The heat of the flame excites electrons in the metal atoms or ions to higher energy levels.
- The excited electrons fall back to lower levels and emit light with energies equal to the gaps.
- Some of those emission lines are in the visible range, and together they give the flame its colour.
Each metal has its own energy levels, so its own colour. A flame test is a simple emission analysis.
| Ion | Flame colour |
|---|---|
| Crimson (deep red) | |
| Intense yellow-orange | |
| Lilac (pale purple) | |
| Orange-red (brick red) | |
| Scarlet red | |
| Pale (apple) green | |
| Blue-green |
Carrying out a flame test
Clean the wire
Dip a nichrome or platinum wire loop in dilute hydrochloric acid and hold it in a hot blue Bunsen flame until it gives no colour. Wear safety glasses.
Load the sample
Dip the clean loop in the acid, then into the solid sample (or a solution of it).
Observe
Hold the loop at the edge of the hot blue flame and record the colour. A hand-held spectroscope shows the individual lines.
Clean between samples
Repeat the cleaning step so no ions carry over to the next test.
Each flame colour comes from a set of emission lines. Compare the unknown's lines with the references to identify it.
A white solid gives a lilac flame when viewed through blue cobalt glass. Which ion does it contain?
Common mistake
Limits of the flame test
- Sodium is a common impurity and its intense yellow can mask other colours.
- Some colours are hard to tell apart (lithium and strontium are both red), and the observer's eye is subjective.
- Many metal ions give no visible colour (for example and ).
- It is qualitative: it tells you which metal is there, not how much.
A spectroscope (looking at the lines, not just the overall colour) or AAS overcomes most of these.
Key idea
Atomic absorption spectroscopy (AAS)
AAS measures how much of one metal is in a sample, down to parts per million or less. Australian scientist Alan Walsh developed it at CSIRO in the 1950s.
- A hollow cathode lamp made of the metal being measured gives out that element's own emission lines.
- The sample solution is sprayed into a flame, which turns it into free ground-state atoms.
- The lamp's light passes through the flame. Atoms of that element absorb exactly those wavelengths, as their electrons jump to higher levels. Other elements' atoms cannot absorb them, so AAS is very selective.
- A detector measures the absorbance: how much of the light was absorbed. More atoms in the flame absorb more light, so absorbance is proportional to concentration (within a working range).
To turn absorbance into concentration, you first measure a series of standard solutions of known concentration, including a blank (the solvent with none of the metal) to zero the instrument. A graph of absorbance against concentration, the calibration curve, should be a straight line. The unknown's absorbance is then read across to the line and down to its concentration.
Worked example
Copper in water from an AAS calibration curve
A student analyses the copper in a sample of water from a hot water system by AAS. The standards gave these absorbances:
| () | 0.0 | 2.0 | 4.0 | 6.0 | 8.0 | 10.0 |
|---|---|---|---|---|---|---|
| Absorbance | 0.000 | 0.105 | 0.208 | 0.316 | 0.419 | 0.522 |
The water sample had an absorbance of 0.310. Determine the copper concentration of the water, and comment on whether it meets the Australian drinking water guideline of .
Calibration curve for copper. The dashed lines read the unknown's absorbance across to the line of best fit and down to the concentration.
Step 1: Check the calibration is linear
The standards lie very close to a straight line through the origin (the blank reads 0.000), so absorbance is proportional to concentration over this range.
Step 2: Find the gradient of the line of best fit
From the line of best fit, gradient (a least-squares fit through the origin gives 0.0523).
The line of best fit uses all six standards, so it averages out small random errors in any one of them.
Step 3: Read off the unknown
. Reading the graph gives about 5.9.
Step 4: Check it is interpolated
0.310 lies between the 4.0 and 6.0 standards, inside the calibrated range, so the reading is reliable.
Step 5: Compare with the guideline
is about three times the guideline, so this water is not suitable for drinking, possibly because of corroding copper pipes.
Try it yourself first, then check the working one step at a time.
Answer
The water contains about of copper ions, roughly three times the guideline value, so it does not meet the guideline.
Method
Plot absorbance against concentration for the standards and draw the line of best fit.
Read the unknown's absorbance across to the line and down, or divide by the gradient.
Check the reading lies within the range of the standards; apply any dilution factor.
Watch out for
Joining the points dot-to-dot instead of using a line of best fit.
Reading concentration on the wrong axis.
Standards of 0–4 mg L⁻¹ lead. A 10.0 mL sample of waste water from a mine site was diluted to 50.0 mL before measurement. Drag the unknown's absorbance to see how the reading changes.
A 20.0 mL sample of mine waste water was diluted to 250.0 mL before AAS analysis. The calibration curve gave a zinc concentration of 3.20 mg L⁻¹ for the diluted solution. Calculate the zinc concentration in the original waste water, in to three significant figures.
Exam tip
Evaluating AAS results
- Interpolate, don't extrapolate. If the unknown's absorbance is above the highest standard, the line may no longer be straight: dilute the sample and measure again.
- Use the line of best fit, not two neighbouring points; it averages out random error in the standards.
- Run a blank and replicates of the unknown; report the mean with its uncertainty, .
- Systematic errors come from inaccurately prepared standards, a dirty burner, or other substances in the sample that change how well the metal atomises.
Summary
Flame tests: heat excites electrons; they fall back and emit light of characteristic colours. Qualitative only, and easily masked by sodium.
AAS: ground-state atoms in a flame absorb their own element's wavelengths from a lamp of that element; absorbance is proportional to concentration.
A calibration curve of standards (with a blank) is read at the unknown's absorbance; use the line of best fit and stay within the range of the standards.
Multiply by the dilution factor (final volume ÷ sample volume) to get the original concentration.
Finished the lesson?
Try the 2 quick checks above and the lesson completes itself, or mark it complete now.