Lesson 8: Flame tests and atomic absorption spectroscopy
Copper in water from an AAS calibration curve
SkillAtomic absorption spectroscopy
Worked example
A student analyses the copper in a sample of water from a hot water system by AAS. The standards gave these absorbances:
| () | 0.0 | 2.0 | 4.0 | 6.0 | 8.0 | 10.0 |
|---|---|---|---|---|---|---|
| Absorbance | 0.000 | 0.105 | 0.208 | 0.316 | 0.419 | 0.522 |
The water sample had an absorbance of 0.310. Determine the copper concentration of the water, and comment on whether it meets the Australian drinking water guideline of .
Step 1: Check the calibration is linear
The standards lie very close to a straight line through the origin (the blank reads 0.000), so absorbance is proportional to concentration over this range.
Step 2: Find the gradient of the line of best fit
From the line of best fit, gradient (a least-squares fit through the origin gives 0.0523).
The line of best fit uses all six standards, so it averages out small random errors in any one of them.
Step 3: Read off the unknown
. Reading the graph gives about 5.9.
Step 4: Check it is interpolated
0.310 lies between the 4.0 and 6.0 standards, inside the calibrated range, so the reading is reliable.
Step 5: Compare with the guideline
is about three times the guideline, so this water is not suitable for drinking, possibly because of corroding copper pipes.
Answer
The water contains about of copper ions, roughly three times the guideline value, so it does not meet the guideline.
Try a similar one
A fresh question of the same kind, checked as you answer, with hints and a full solution.
Calibration curve for copper. The dashed lines read the unknown's absorbance across to the line of best fit and down to the concentration.
Method
Plot absorbance against concentration for the standards and draw the line of best fit.
Read the unknown's absorbance across to the line and down, or divide by the gradient.
Check the reading lies within the range of the standards; apply any dilution factor.
Watch out for
Joining the points dot-to-dot instead of using a line of best fit.
Reading concentration on the wrong axis.