Lesson 8: Flame tests and atomic absorption spectroscopy
Evaluating an AAS analysis of an iron-fortified drink
SkillAtomic absorption spectroscopy
Worked example
An iron-fortified drink is labelled as containing of iron. A student diluted 25.0 mL of it to 100.0 mL and measured the absorbance of the diluted solution three times: 0.224, 0.229 and 0.225. The calibration line for the iron standards passes through the origin with a gradient of .
(a) Calculate the iron concentration of the drink.
(b) Determine the absolute and percentage uncertainty in the mean absorbance.
(c) Evaluate the label's claim.
Step 1: Mean absorbance
Step 2: Concentration of the diluted solution
Step 3: Apply the dilution factor
Dilution factor , so the drink contains .
Step 4: Uncertainty of the mean
percentage uncertainty , so the concentration is .
The concentration is proportional to the absorbance, so it carries the same percentage uncertainty.
Step 5: Percentage error against the label
percentage error
Step 6: Evaluate
The measured value, , does not reach 21.0, and the 4.8 % difference is about four times the 1.1 % uncertainty, so random error in the readings does not explain it. Either the drink contains slightly less iron than claimed, or there is a systematic error: inaccurate standards, a volumetric error in the dilution, or other substances in the drink affecting atomisation. The claim is not supported by this analysis, but the difference is small; repeating with freshly prepared standards and a certified reference sample would test it.
Answer
(a) . (b) (1.1 %). (c) The result is 4.8 % below the label, more than the 1.1 % uncertainty, so the claim is not supported; the difference suggests a systematic error or a slightly lower iron content, to be checked with new standards and a reference sample.
Try a similar one
A fresh question of the same kind, checked as you answer, with hints and a full solution.
Method
Average the replicates and find the uncertainty of the mean.
Convert absorbance to concentration with the calibration line, then apply the dilution factor.
Compare the percentage error with the percentage uncertainty to judge the claim, and name likely sources of error.
Watch out for
Forgetting the dilution factor and reporting 5.01 mg/L.
"Human error" as the only explanation. Name a specific source and say which way it would push the result.